Characterization of a pressurized C5-C16 hydrocarbon gas calibration standard for air analysis
Detlev Helmig1, Tobias Revermann, Bradley Hall
1Institute of Arctic and Alpine Research (INSTAAR), University of Colorado, Boulder, Colorado 80309-0450, USA. Detlev.Helmig@instaar.colorado.edu
Abstract:
A compressed gas standard containing parts-per-billion (ppb) amounts of the volatile hydrocarbons methylpentadiene (isoprene, 540 ppb) and isooctane (259 ppb) and a series of less volatile C12-C16 n-alkanes (n-dodecane, 349 ppb; n-tridecane, 340 ppb; n-tetradecane, 202 ppb; n-pentadecane, 271 ppb; and n-hexadecane, 308 ppb) was prepared by a one-step (no further gas dilution) microgravimetric method. The gravimetric mixing ratios were confirmed by referencing to a capillary diffusion method. The cylinder was heated to 75 degrees C to minimize condensation losses of analytes to the cylinder walls. Mixing ratios were monitored over a 2.5-year period. Some initial analyte losses (approximately 3 to 20%) were observed for the heavier C14-C16 n-alkanes. Subsequently, analyte loss rates were found to be in the range of <1.0% per year for n-dodecane to n-hexadecane. The developed guidelines for preparation, storage, and retrieval of these semivolatile analytes enable the use of compressed gas standards for calibration and method development purposes in the environmental gas-phase analysis of these and related compounds.
Related Concept Videos
Applications of the Ideal Gas Law: Molar Mass, Density, and Volume
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Standard Enthalpy of Formation
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Ideal Gas Equation