Related Experiment Video
Updated: Aug 21, 2026

Direct Microbial Identification using An Automated Microbial Identification System to Facilitate the EUCAST RAST Method Without Mass Spectrometry
Published on: May 24, 2024
Evaluation of API ID 32C and VITEK-2 to identify Candida dubliniensis
Carmen-Delia Cárdenes-Perera1, Alvaro Torres-Lana, Rocío Alonso-Vargas
1Area de Medicina Preventiva y Salud Pública, Facultad de Medicina, Universidad de La Laguna, La Laguna, Spain.
Abstract:
We have compared two diagnostic systems, API-ID32C (bioMerieux) and VITEK-2ID-YST (bioMerieux), in their ability to diagnose 50 clinical isolates of Candida dubliniensis. API identified 48 isolates and VITEK-2 identified 33 of the 50 isolates. Lactic acid assimilation showed highly dispersed results with API, being of special importance, since this test is one of the four that makes it possible to differentiate C. dubliniensis from C. albicans. The detection of one enzyme, Phosphate-4MU, was always positive with VITEK-2 when it should have been negative. Overall, API ID 32C obtained better results than VITEK-2. However, the latter is simpler to use and has a greater database.