A Simple Dewar/Cryostat for Thermally Equilibrating Samples at Known Temperatures for Accurate Cryogenic Luminescence Measurements
Published on: July 19, 2016
Development of a stabilized low temperature infrared absorption cell for use in low temperature and collisional
A Valentin1, A Henry, C Claveau
1Laboratoire de Physique Moléculaire et Applications, Université P. et M. Curie et CNRS, 4 Place Jussieu, Tour 13, Case 76, F-75252 Paris Cedex 05, France.
Abstract:
We have constructed a stabilized low temperature infrared absorption cell cooled by an open cycle refrigerator, which can run with liquid nitrogen from 250 to 80K or with liquid helium from 80K to a few kelvin. Several CO infrared spectra were recorded at low temperature using a tunable diode laser spectrometer. These spectra were analyzed taking into account the detailed effects of collisions on the line profile when the pressure increases. We also recorded spectra at very low pressure to accurately model the diode laser emission. Spectra of the R(2) line in the fundamental band of 13CO cooled by collisions with helium buffer gas at 10.5K and at pressures near 1 Torr have been recorded. The He-pressure broadening parameter (gamma(0) = 0.3 cm(-1) atm(-1)) has been derived from the simultaneous analysis of four spectra at different pressures.