Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Gradient field switching as a source for artifacts in MR imaging of metallic stents
Avshalom Shenhav1, Haim Azhari
1Department of Biomedical Engineering, Technion, Israel Institute of Technology, Haifa.
Abstract:
Metallic implants, such as stents, have long been a concern in magnetic resonance imaging (MRI). In addition to safety issues, they are commonly associated with image artifacts. The mechanisms of radiofrequency- (RF) and susceptibility-induced artifacts have been thoroughly investigated. However, gradient-switching-induced artifacts have not been analyzed. In this study it was demonstrated that gradient switching may be a source of artifacts in metallic stent MR imaging. These artifacts differ from those caused by the RF pulse. A theoretical explanation is provided as well.