A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Upper limit of vulnerability determination during implantable cardioverter-defibrillator placement to minimize
Michael Glikson1, Osnat T Gurevitz, Jane M Trusty
1Sheba Medical Center and Tel Aviv University, Tel Hashomer, Israel.
Abstract:
The defibrillation threshold (DFT) and upper limit of vulnerability (ULV) were determined using step-down protocols in 50 patients who underwent implantable cardioverter-defibrillator placement or testing. The sensitivity and specificity of each ULV energy level was assessed for detecting an increased DFT, correlation of the DFT and ULV, and optimal shock timing for ULV determination. A ULV <10 or 11 J (failure to induce ventricular fibrillation with 10- to 11-J shocks) was 100% predictive of an acceptable DFT and may be sufficient to exclude unacceptable DFTs in 60% of implantable cardioverter-defibrillator recipients. All 4 shocks used to scan the peak of the T wave during ULV testing were necessary for accurate ULV determination.