Insights

Metasul metal-on-metal hip replacements show clinical success with typical mechanical complications. Evidence supports continued use for patients without kidney issues.

Area of Science:

  • Orthopedic surgery
  • Biomaterials science

Background:

  • Metasul metal-on-metal articulations have been utilized in approximately 300,000 total hip replacements over 15 years.
  • Clinical success has been demonstrated through Harris hip scores and patient self-assessment in previous studies.

Purpose of the Study:

  • To evaluate the clinical outcomes and complications associated with Metasul articulations in total hip replacements.
  • To compare the performance of Metasul articulations against ceramic-on-polyethylene bearings.

Main Methods:

  • Three clinical studies involving Metasul articulations were reviewed.
  • A randomized study compared Metasul to ceramic-on-polyethylene bearings.
  • Patient outcomes were assessed using Harris hip scores and self-assessment.

Main Results:

  • Mechanical complications included two cup loosenings and 24 dislocations in 582 patients (619 hips), a 3.8% rate.
  • No significant differences in clinical results or complications were observed between Metasul and ceramic-on-polyethylene groups in the randomized study.
  • Literature reports low annual linear wear rates and no adverse effects from elevated cobalt ion levels for Metasul implants.

Conclusions:

  • The scientific evidence supports the continued use of Metasul articulations in total hip replacements.
  • Metasul implants are recommended for patients without compromised renal function.