Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Fixation strength of an all-metal acetabular component cemented into an acetabular shell: a biomechanical analysis
Edward Ebramzadeh1, Paul E Beaulé, Joelle Lee Culwell
1Biomechanics Laboratory, J Vernon Luck, Sr, MD Orthopaedic Research Center, Los Angeles Orthopaedic Hospital/UCLA, Los Angeles, California 90007-2697, USA.
Abstract:
When an acetabular shell is well fixed but the locking mechanism is compromised, cementing a new liner into the existing acetabular component can provide a simple revision solution. The pull-out and torsional fixation strength of cobalt chromium (CoCr) alloy acetabular components cemented into 3 sizes of titanium alloy acetabular components was tested under conditions of 0 or 2 mm cement mantle at the dome and 2 mm cement with the cemented component in 20 degrees of version with respect to the shell. The lowest mean tensile load to failure was 1500 N, and occurred with no cement at the dome of the 54 mm shell, whereas the greatest load was with 2 mm cement thickness. Smallest mean torque to failure was 43 Nm. Version angle did not substantially change failure load. Although not cyclically loaded, fixation strength of metal liners cemented into shells was comparable to that of commonly used locking mechanisms, suggesting sufficient strength for clinical application.
Related Concept Videos
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in ASTM C...
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...