Acute tensile strength analysis of collagen solder for mesh fixation to the peritoneal surface
R J Lanzafame1, B A Soltz, I Stadler
1Laser Center, Rochester General Hospital, 1445 Portland Avenue # 202, Rochester, NY 14621-3095, USA. ray.lanzafame@viahealth.org
Background:
In this study, we assessed the feasibility of laser-assisted tissue welding as a means of fixing mesh prostheses to the peritoneum. We then tested the initial tensile strength of the bonds.
Methods:
Fresh porcine peritoneal coupons were lap-joint bonded with laser-activated solder. Anesthetized New Zealand white rabbits and Yorkshire pigs also underwent laparotomy. Vicryl mesh (2.0 x 1.0 cm) was attached to the peritoneum using a laser system (1.43 mu, 2.5 W, 60 degrees C), solder formulations and configurations, and a 1 cm2 bond area. Control segments were affixed with 4.8-mm staples. After the animals were killed, the segments were excised en bloc. Tensile strength assessment was conducted by measuring peak force breaking strength.
Results:
The strength of the solder bonds were similar in all groups (range, 261.5 +/- 170.3-465.3 +/- 288.2 g/cm2) and were not statistically different from the controls (215.8 +/- 117.8 g/cm2).
Conclusions:
These values are similar to the 200-500 g/cm2 acute strengths reported for sutured or stapled peritoneal closure. Mesh fixation by solder is feasible, and further development of this technology is warranted.