External defibrillation in the left lateral position--a comparison of manual paddles with self-adhesive pads
Tamsin E L Dodd1, Charles D Deakin, Graham W Petley
1Shackleton Department of Anaesthetics, Southampton University Hospital NHS Trust, Tremona Road, Southampton SO166YD, UK.
Introduction:
Firm paddle force during defibrillation lowers transthoracic impedance (TTI) and increases transmyocardial current, increasing the chances of successful cardioversion. Current protocols recommend that if defibrillation using the anterior-apical (AA) paddle position fails, the anterior-posterior (AP) position should be used. This generally requires the patient to be placed in the left lateral position with the operator leaning over the patient. Avoiding physical contact with the patient during defibrillation subjectively makes application of firm paddle force difficult in the AP position. We compared TTI between the AA and AP positions and between manual paddles and self-adhesive pads to establish if the AP position precludes firm paddle force and to compare TTI between paddles and self-adhesive pads.
Methods:
Twenty-one consecutive patients undergoing elective cardioversion (age 39-82) were studied. TTI was measured between pairs of manually held paddles and self-adhesive pads using AA placement with the patient supine, and AP placement with the patient left lateral position.
Results:
Mean TTI using the AP electrode position was lower using manual paddles (66.5 Omega; 95% CI 60.2-72.9 Omega) than that using self-adhesive pads (92.1 Omega; 95% CI 81.5-102.7 Omega; 95% CI between the mean =15.8-35.5 Omega; P <0.0001). TTI was significantly less using the manual paddles compared with self-adhesive pads in both AA and AP positions (P <0.0001).
Conclusion:
Despite the subjective difficulties of defibrillating patients in the AP position whilst leaning over them, use of manual paddles achieves a lower TTI than that achieved with self-adhesive pads.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...