Gas chromatography/ion trap mass spectrometry applied for the determination of polybrominated diphenyl ethers in soil
Dongli Wang1, Zongwei Cai, Guibin Jiang
1Department of Chemistry, Hong Kong Baptist University, Hong Kong SAR, China.
Abstract:
A gas chromatography/ion trap mass spectrometry (GC/ITMS) method was developed for the determination of polybrominated diphenyl ethers (PBDEs). ITMS parameters were optimized in order to achieve the best sensitivity for the PBDE analysis. Tandem mass spectrometry, along with an isotope dilution internal standard method, was used for the quantitation. Chromatographic windows were developed for mono- to hepta-BDEs, depending on the retention times when a 30-m GC column was used. A different 15-m column was used to analyze deca-BDE. Environmental soil samples collected from an electronic waste recycling site were prepared by using Soxhlet extraction and column chromatographic cleanup. Average recoveries of 61-118% were obtained for the 13C-labeled PBDE internal standards spiked in the samples prior to sample preparation. The accuracy represented by relative analytical errors was -24% to 18%, and the precision (relative standard deviation) was 11-26% (n=8). The method detection limits ranged from 0.013-0.25 ng/g for the PBDEs in soil.
More Related Videos
04:17Analysis of Organochlorine Pesticides in a Soil Sample by a Modified QuEChERS Approach Using Ammonium Formate
Published on: January 20, 2023
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...