Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
A fast, convenient and rugged procedure for characterizing the selectivity of alkyl-silica columns
L R Snyder1, A Maule, A Heebsh
1LC Resources Inc, Orinda, CA 94563, USA. snyder0036@comcast.net
Abstract:
Previous work has shown that the selectivity of reversed-phase columns for HPLC can be described by means of five column parameters: H (hydrophobicity), S* (steric resistance), A (hydrogen-bond acidity), B (hydrogen-bond basicity) and C (cation-exchange capacity). Values of H, S*, etc. can be determined by carrying out retention measurements for 18 test solutes under standardized conditions. The reproducibility of the latter procedure has been evaluated by comparison testing in four different laboratories and found acceptable. An alternative 10-solute test procedure which is more reproducible and convenient (but somewhat less accurate), requires only 2-3 h per column.