A Stable Phantom Material for Optical and Acoustic Imaging
Published on: June 16, 2023
Thickness sensitivity of ultrasound velocity in long bone phantoms
Petro Moilanen1, Vantte Kilappa, Patrick H F Nicholson
1Department of Health Sciences, University of Jyväskylä, Finland. pemoilan@cc.jyu.fi
Abstract:
One approach to bone disease diagnosis such as osteoporosis is to measure the velocity of ultrasound propagating axially along long bones. In this study, the variation in velocity as a function of radial position was assessed using two polyvinyl chloride (PVC) bone phantoms with cross-sectional geometry similar to the human tibia but differing in medullary cavity diameter. Two ultrasonometers were used: these were a commercial device operating at a relatively high frequency (HF) of 1.25 MHz and a prototype low frequency (LF) device operating at approximately 200 kHz. The LF measurements showed a larger variation with radial position, with changes in velocity of up to 20% occurring around the phantom compared with changes of only 4% at most for HF. The LF velocity correlated strongly with local thickness (r(2) = 0.81) but HF velocity did not. The results demonstrate that LF measurements have a greatly enhanced thickness sensitivity. Using LF, it may therefore be possible to assess bone thickness as a function of radial position and hence to determine the distribution of bone around the long axis.