Lack of equivalence between central and mixed venous oxygen saturation
Lakhmir S Chawla1, Hasan Zia, Guillermo Gutierrez
1The George Washington University MFA, 2150 Pennsylvania Ave NW, Suite 5-404, Washington, DC 20037, USA.
Study Objective:
We compared paired samples of central venous O(2) saturation (Scvo(2)) and mixed venous O(2) saturation (Svo(2)) to test the hypothesis that Scvo(2) is equivalent to Svo(2). We also compared O(2) consumption (Vo(2)) computed with Scvo(2) (Vo(2)cv) to that computed with Svo(2) (Vo(2)v).
Design:
Prospective, sequential, observational study.
Setting:
Combined medical-surgical ICU.
Patients:
Fifty-three individuals > 18 years of age of either sex who required a pulmonary artery catheter (PAC) to guide fluid therapy. Subjects were identified as postsurgical (32 patients) or medical (21 patients) according to their ICU admission diagnosis.
Interventions:
A PAC was inserted through the internal jugular or subclavian veins. Care was taken to place the PAC proximal port approximately 3 cm above the tricuspid valve. Blood samples were drawn from the proximal and distal ports in random order. An arterial blood sample also was drawn.
Measurements:
Cardiac output in triplicate, systemic pressure, and central pressure. We analyzed blood samples for hemoglobin concentration and O(2) saturation (So(2)). Data were compared by correlation analysis and by the method of Bland and Altman.
Results:
Svo(2) was consistently lower than Scvo(2) (p < 0.0001), with a mean (+/-SD) bias of -5.2 +/- 5.1%. Similar differences in Scvo(2) and Svo(2) were present within each subgroup (p < 0.001). A lower Svo(2) resulted in Vo(2)v values that were higher than the Vo(2)cv values for all patients in the study (mean Vo(2)v, 236.7 +/- 103.4 mL/min; mean Vo(2)cv, 191.1 +/- 84.0 mL/min; p < 0.001) as well as for patients within each subgroup (p < 0.001).
Conclusions:
Measurements of Scvo(2) and Svo(2) were not equivalent in this sample of critically ill patients. Moreover, substituting Scvo(2) for Svo(2) in the calculation of Vo(2) produced unacceptably large errors. The decrease in So(2) between Scvo(2) to Svo(2) may result from the mixing of atrial and coronary sinus blood. As such, this difference may be a marker of myocardial O(2) consumption.
More Related Videos
09:04Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Transport in the Blood
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...