Related Experiment Video
Updated: Aug 20, 2026

Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization
Published on: August 6, 2018
Orthogonal trap time-of-flight mass spectrometer using a collisional damping chamber
Yuichiro Hashimoto1, Izumi Waki, Kiyomi Yoshinari
1Central Research Laboratory, Hitachi, Ltd., Tokyo, Japan. h-yuichi@crl.hitachi.co.jp
Abstract:
We report a new hybrid mass spectrometer, which is a combination of a quadrupole ion trap and an orthogonal time-of-flight (TOF) mass spectrometer. This new configuration consists of a collisional-damping chamber (CDC) inserted between an MSn-capable ion trap and a high-mass-accuracy orthogonal-TOF mass spectrometer. Because the CDC converted an ion packet into an energy-focused and quasi-continuous beam, a high mass resolution of over 10,000 and a high mass accuracy of better than 3 ppm were achieved. Moreover, the ratio of the maximum detectable m/z to the minimum detectable m/z, which is referred to here as the mass window, was improved to more than 10.
Related Concept Videos
Mass Analyzers: Common Types
Tandem Mass Spectrometry
Mass Analyzers: Overview
MALDI-TOF Mass Spectrometry
Mass Spectrometers
Peptide Identification Using Tandem Mass Spectrometry