Objective measures like electrically evoked compound action potential (ECAP) and auditory brain stem response (EABR) help determine cochlear implant (CI) stimulation levels in children. These non-behavioral tests, including the electrically evoked stapedius reflex (ESR), are feasible in the operating room and post-surgery.

Area of Science:

  • Audiology
  • Biomedical Engineering
  • Neuroscience

Background:

  • Cochlear implants (CIs) require immediate audible and comfortable stimulation post-activation.
  • Objective measures are crucial for setting initial CI parameters in children.

Purpose of the Study:

  • To assess the feasibility of recording ECAP, EABR, and ESR in children at implantation and over the first year.
  • To evaluate changes in ECAP, EABR, and ESR thresholds and their relation to behavioral measures over time.
  • To determine the predictive value of ECAP, EABR, and ESR for behavioral thresholds and loudness in pediatric CI users.

Main Methods:

  • Recorded ECAP, EABR, and ESR responses in 68 children over one year.
  • Utilized electrical pulses from three different electrodes along the CI array.
  • Performed visual inspection of stapedius reflex (V-ESR) in an additional 20 children during surgery.

Main Results:

  • ECAP and EABR were successfully recorded in over 84% of electrodes and 89% of children, both in the operating room and post-surgery.
  • ESRs were recorded in over 65% of electrodes and 67% of children by 3 months, but less reliably in the operating room.
  • ECAP and EABR thresholds remained stable, while ESR thresholds increased; behavioral thresholds decreased and maximum levels increased over time.

Conclusions:

  • Non-behavioral measures (ECAP, EABR, ESR) aid in determining CI stimulation levels, especially in young children.
  • These measures are feasible in the operating room and can be repeated post-surgery.
  • ECAP and EABR thresholds can predict stimulation levels, with adjustments for electrode position and child's age, while ESR best predicts maximum levels.