Related Experiment Video
Updated: Aug 20, 2026

A Novel Approach for the Administration of Medications and Fluids in Emergency Scenarios and Settings
Published on: November 9, 2016
Interaction between infusion equipment resulting in drug overdose in a critically ill patient
1Anaesthetics Department, Victoria Infirmary, Langside, Glasgow, G42 9TY, Scotland.
Abstract:
A critically ill septic patient on haemofiltration for acute renal failure suffered sudden circulatory and respiratory collapse. The cause of the collapse was traced to an interaction between mechanical devices (syringe driver infusion pumps and haemofiltration equipment) connected to two central venous catheters.