Related Experiment Video
Updated: Aug 14, 2026

T-wave Ion Mobility-mass Spectrometry: Basic Experimental Procedures for Protein Complex Analysis
Published on: July 31, 2010
Systematic comparison of a two-dimensional ion trap and a three-dimensional ion trap mass spectrometer in proteomics
Viveka Mayya1, Karim Rezaul, Yu-Sheng Cong
1Center for Vascular Biology, Department of Cell Biology, University of Connecticut Health Center, 263 Farmington Avenue, Farmington, CT 06030, USA.
Abstract:
The utility and advantages of the recently introduced two-dimensional quadrupole ion trap mass spectrometer in proteomics over the traditional three-dimensional ion trap mass spectrometer have not been systematically characterized. Here we rigorously compared the performance of these two platforms by using over 100,000 tandem mass spectra acquired with identical complex peptide mixtures and acquisition parameters. Specifically we compared four factors that are critical for a successful proteomic study: 1) the number of proteins identified, 2) sequence coverage or the number of peptides identified for every protein, 3) the data base matching SEQUEST X(corr) and S(p) score, and 4) the quality of the fragment ion series of peptides. We found a 4-6-fold increase in the number of peptides and proteins identified on the two-dimensional ion trap mass spectrometer as a direct result of improvement in all the other parameters examined. Interestingly more than 70% of the doubly and triply charged peptides, but not the singly charged peptides, showed better quality of fragmentation spectra on the two-dimensional ion trap. These results highlight specific advantages of the two-dimensional ion trap over the conventional three-dimensional ion traps for protein identification in proteomic experiments.
More Related Videos
10:37Deep Proteome Profiling by Isobaric Labeling, Extensive Liquid Chromatography, Mass Spectrometry, and Software-assisted Quantification
Published on: November 15, 2017
07:33Analyzing Protein Architectures and Protein-Ligand Complexes by Integrative Structural Mass Spectrometry
Published on: October 15, 2018
Related Concept Videos
Mass Spectrometers
MALDI-TOF Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Analyzers: Overview
Mass Analyzers: Common Types
Tandem Mass Spectrometry