Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
SEM observations of nickel-titanium rotary endodontic instruments that fractured during clinical Use
Satish B Alapati1, William A Brantley, Timothy A Svec
1Section of Oral Biology, College of Dentistry, The Ohio State University, Columbus, OH 43218-2357, USA.
Abstract:
Numerous discarded ProFile GT, ProFile, and ProTaper nickel-titanium rotary instruments obtained from two graduate endodontic clinics were examined by scanning electron microscopy. These instruments had an unknown history of clinical use and had fractured or experienced considerable permanent torsional deformation without complete separation. The failure processes generally exhibited substantial ductile character, evidenced by a dimpled rupture fracture surface. Crack propagation at grain boundaries and cleavage surfaces indicative of transgranular fracture were observed for some specimens. It appeared that oxide particles from the manufacturing process served as nucleating sites for the microvoids, leading to dimpled rupture. A previously unreported fracture mode also was observed, in which crack propagation, approximately parallel to the local flute orientation, connected pitted regions on the surface. Combining present and previous scanning electron microscopy observations of clinically failed instruments, suggestions are offered for improving their fracture resistance.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
06:48Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025