Validation of a new non-invasive portable tonometer for determining arterial pressure wave and pulse wave velocity:
Paolo Salvi1, Giuseppe Lio, Carlos Labat
1Department of Internal Medicine, Bufalini Hospital, Cesena, Italy. psalvi.md@libero.it
Objective:
To validate a new, small portable tonometer (PulsePen) that is able to assess carotid artery pressure and to measure pulse wave velocity (PWV) non-invasively. Its software provides absolute arterial pressure values, an assessment of arterial pulse wave contours, an estimation of reflection waves and measurements of PWV.
Design And Methods:
Two validation studies were carried out. The aim of the first study was to compare arterial pressure values and pulse wave contours recorded in the carotid artery using the PulsePen versus intra-arterial simultaneous measurements in 10 patients undergoing cardiac catheterization. The pulse wave contour was assessed using Fourier analysis. The comparison between the two methods showed no difference in arterial pressure wave spectral moduli from harmonics 1 to 6. The second study compared PWV measurements taken with the PulsePen (one tonometer) and measurements performed with two Millar tonometers in 68 subjects (32 men, 36 women). PulsePen measurements were realized as two consecutive measurements in the carotid and femoral arteries, both synchronized by electrocardiogram. The pulse wave transit time was calculated as the difference between the time delay of the femoral pulse wave and the carotid pulse wave in relation to the R wave of the electrocardiogram. These measurements were compared with PWV obtained by simultaneous carotid and femoral measurements with the two Millar tonometers. No difference between the two methods was found, with a variation coefficient of 7.7%. The variation coefficients of the inter-observer and intra-observer reproducibility for the PulsePen were 7.9 and 7.2%, respectively.
Conclusions:
These results show that the PulsePen enables an easy and reliable evaluation of central arterial pressure and stiffness in clinical ambulatory practice, especially in high-risk patients in whom arterial stiffness has been shown to be a significant indicator of morbidity and mortality.
More Related Videos
05:51Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
06:04Pulse-Wave Velocity, Flow-Mediated Dilation, and Carotid Intima-Media Thickness to Assess Cardiovascular Risk in Population with Metabolic Syndrome
Published on: September 27, 2024
Related Concept Videos
Pulse Assessment Sites
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse
The pulse serves as a clinical indicator...