Related Experiment Video
Updated: Aug 20, 2026

An Experimental Protocol for Studying Mineral Effects on Organic Hydrothermal Transformations
Published on: August 8, 2018
Investigation of retention on bare silica using reversed-phase mobile phases at elevated temperatures
Brian A Bidlingmeyer1, John Henderson
1Agilent Technologies, Little Falls Site, 2850 Centerville Road, Wilmington, DE 19808, USA. brian_bidlingmeyer@agilent.com
Abstract:
The use of unbonded silica as a stationary phase in reversed-phase HPLC is described as a useful alternative to bonded phase columns for polar, lipophilic amines. Using four lipophilic amines, the role of temperature is shown to favorably impact both efficiency and selectivity, which is not universally seen when using bonded phases. As temperature is raised, retention drops on the silica column. The temperature behavior appears to support the hypothesis that retention is dependant upon electrostatic and adsorptive forces.
Related Concept Videos
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Analyte Adsorption and Distribution
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Thin-Layer Chromatography (TLC): Overview
To begin the analysis, a mixture of compounds is spotted on the starting line on the TLC plate using a thin capillary. The bottom of the...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...