Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
Diagnostic accuracy of noninvasive coronary imaging using 16-detector slice spiral computed tomography with 188 ms
Axel Kuettner1, Torsten Beck, Tanja Drosch
1Department for Radiology, Division of Diagnostic Radiology, Eberhard-Karls-University of Tuebingen, Germany.
Objectives:
The aim of our study was to evaluate the diagnostic accuracy of 16-multi-detector spiral computed tomography (MDCT) with 188 ms temporal resolution.
Background:
Because of rapid technical innovations, MDCT coronary imaging has significantly improved in the last five years. Recent results indicate a high diagnostic accuracy, especially in patients with clinical suspicion of coronary artery disease (CAD).
Methods:
A total of 72 consecutive patients (30 women, 42 men, age 64 +/- 10 years) scheduled for invasive coronary angiography (ICA) because of suspected CAD were additionally studied by MDCT (Sensation 16 Speed 4D, Siemens, Forchheim, Germany). Thirty-seven of 72 patients (51%) received an additional beta-blockade before MDCT because of heart rates >65 beats/min. The MDCT scans were analyzed regarding the presence of coronary artery lesions. Results were compared with ICA.
Results:
All 72 scans showed diagnostic image quality (heart rate: 64.1 +/- 9.2 beats/min, calcium mass: 86 +/- 156 mg). Thirteen coronary segments were evaluated in each patient. Sixty-two of 936 (6.6%) segments showed a nondiagnostic image quality. All segments were included in the analysis. A total of 117 relevant lesions (diameter stenosis >50%) were detected using ICA, and 96 of 117 (82%) were detected by MDCT. Sensitivity, specificity, and positive and negative predictive values for the whole study group were as follows: 82%, 98%, 87%, and 97%, respectively. The correct clinical diagnosis of presence or absence of significant CAD was obtained in 65 of 72 (90%) patients. All stenoses were detected by MDCT in 52 of 72 (72%) patients.
Conclusions:
Our results indicate a high diagnostic accuracy of 16-slice MDCT with improved temporal resolution.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System IV: CMRI