Headspace gas chromatographic method for the measurement of difluoroethane in blood
L A Broussard1, A Broussard, T Pittman
1Department of Clinical Laboratory Sciences, LSU Health Sciences Center, 1900 Gravier, New Orleans, LA 70112, USA. lbrous@lsuhsc.edu
Objectives:
To develop a gas chromatographic assay for the analysis of difluoroethane, a volatile substance, in blood and to determine assay characteristics including linearity, limit of quantitation, precision, and specificity.
Setting:
Referral toxicology laboratory
Results:
Difluoroethane, a colorless, odorless, highly flammable gas used as a refrigerant blend component and aerosol propellant, may be abused via inhalation. A headspace gas chromatographic procedure for the identification and quantitation of difluoroethane in blood is presented. A methanolic stock standard prepared from pure gaseous difluoroethane was used to prepare whole blood calibrators. Quantitation of difluoroethane was performed using a six-point calibration curve and an internal standard of 1-propanol. The assay is linear from 0 to 115 mg/L including a low calibrator at 4 mg/L, the limit of quantitation. Within-run coefficients of variation at mean concentrations of 13.8 mg/L and 38.5 mg/L were 5.8% and 6.8% respectively. Between-run coefficients of variation at mean concentrations of 15.9 mg/L and 45.7 mg/L were 13.4% and 9.8% respectively. Several volatile substances were tested as potential interfering compounds with propane having a retention time identical to that of difluoroethane.
Conclusions:
This method requires minimal sample preparation, is rapid and reproducible, can be modified for the quantitation of other volatiles, and could be automated using an automatic sampler/injector system.