Related Experiment Video
Updated: Aug 3, 2026

Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
Stereoselective high-performance liquid chromatographic determination of flurbiprofen in human plasma
G Geisslinger1, S Menzel-Soglowek, O Schuster
1Department of Pharmacology and Toxicology, University of Erlangen-Nürnberg, Germany.
Abstract:
An enantioselective high-performance liquid chromatographic assay for the quantitation of the enantiomers of flurbiprofen in human plasma is described. The procedure involved extraction of flurbiprofen from acidified plasma into hexane-diethyl ether (8:2, v/v). Stereoselective separation was achieved with a prepacked alpha 1-acid glycoprotein column without any derivatization procedure. A second assay using a conventional reversed-phase column to determine racemic flurbiprofen is also described. The detection wavelength was set at 246 nm. The limit of quantification was found to be 50 ng/ml for both assays. The method was demonstrated to be sufficiently sensitive for stereoselective pharmacokinetic studies of flurbiprofen.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Types of Detectors
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
![Radiosynthesis of 1-(2-[18F]Fluoroethyl)-L-Tryptophan using a One-pot, Two-step Protocol](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F63025.jpg&w=3840&q=50)