Related Experiment Video
Updated: Aug 20, 2026

A Microfluidic Chip for the Versatile Chemical Analysis of Single Cells
Published on: October 15, 2013
Microfluidic chip for peptide analysis with an integrated HPLC column, sample enrichment column, and nanoelectrospray
Hongfeng Yin1, Kevin Killeen, Reid Brennen
1Agilent Laboratories, 3500 Deer Creek Road, Palo Alto, California 94304, USA. hongfeng_yin@agilent.com
Abstract:
Current nano-LC/MS systems require the use of an enrichment column, a separation column, a nanospray tip, and the fittings needed to connect these parts together. In this paper, we present a microfabricated approach to nano-LC, which integrates these components on a single LC chip, eliminating the need for conventional LC connections. The chip was fabricated by laminating polyimide films with laser-ablated channels, ports, and frit structures. The enrichment and separation columns were packed using conventional reversed-phase chromatography particles. A face-seal rotary valve provided a means for switching between sample loading and separation configurations with minimum dead and delay volumes while allowing high-pressure operation. The LC chip and valve assembly were mounted within a custom electrospray source on an ion-trap mass spectrometer. The overall system performance was demonstrated through reversed-phase gradient separations of tryptic protein digests at flow rates between 100 and 400 nL/min. Microfluidic integration of the nano-LC components enabled separations with subfemtomole detection sensitivity, minimal carryover, and robust and stable electrospray throughout the LC solvent gradient.