Choice of oximeter affects apnea-hypopnea index
Subooha Zafar1, Indu Ayappa, Robert G Norman
1New York University School of Medicine, New York, NY, USA.
Study Objectives:
Current Medicare guidelines include an apnea-hypopnea index (AHI) > or = 15 events per hour, in which all hypopneas must be associated with 4% desaturation, to qualify for reimbursement for therapy with continuous positive airway pressure (CPAP). The present data demonstrate the effect of pulse oximeter differences on AHI.
Design:
Prospective study, blinded analysis.
Setting:
Academic sleep disorder center.
Patients:
One hundred thirteen consecutive patients (84 men and 29 women) undergoing diagnostic sleep studies and being evaluated for CPAP based on the Medicare indications for reimbursement.
Interventions:
Patients had two of four commonly used oximeters with signal averaging times of 4 to 6 s placed on different digits of the same hand during nocturnal polysomnography.
Measurements And Results:
Apneas and candidate hypopneas (amplitude reduction, > 30%) were scored from the nasal cannula airflow signal without reference to oximetry. Candidate hypopneas then were reclassified as hypopneas by each oximeter if they were associated with a 4% desaturation. Although the use of three oximeters resulted in a similar AHI (bias, < 1 event per hour), the fourth oximeter showed an overall increase in AHI of 3.7 events per hour. This caused 7 of 113 patients to have an AHI of > or = 15 events per hour (meeting the Medicare criteria for treatment) by one oximeter but not when a different oximeter was used. More importantly, when our analysis was limited to those patients whose number of candidate hypopneas made them susceptible to the threshold value of 15 events per hour, 7 of 35 patients who did not meet the Medicare AHI standard for treatment by one oximeter were reclassified when a different oximeter was used.
Conclusion:
In the present study, oximeter choice affected whether the AHI reached the critical cutoff of 15 events per hour, particularly in those with disease severity that was neither very mild nor very severe. As oximetry is not a technique that produces a generic result, there are significant limitations to basing the definition of hypopnea on a fixed percentage of desaturation in determining the eligibility for CPAP therapy.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...