Related Experiment Video
Updated: Aug 20, 2026

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
[Are angle-fixed implants with elastic properties advantageous for the internal fixation of proximal humerus
M H Hessmann1, W Sternstein, D Mehler
1Klinik und Poliklinik für Unfallchirurgie der Johannes Gutenberg-Universität, Mainz. Hessmann@unfall.klinik.uni-mainz.de
Abstract:
There is a recent interest for the use of angle-fixed plates in the management of proximal humerus fractures. Rigid implants might be associated with an increased risk of cutting-out. In order to analyse the potential beneficial effects of the implant elasticity on fracture fixation, the biomechanical properties of a rigid and an elastic angle-fixed plating system were assessed in an experimental study. An unstable fracture of the surgical neck was created in 8 pairs of human humeri. Specimens were subjected to axial loading and torque. Stiffness, subsidence and load to failure were assessed. The implant with elastic properties was characterized by a lower torsional stiffness and a higher subsidence during axial loading and torque. This implant failed at lower loads than the rigid implant did. Elastic implant properties of angle-fixed plates have shown not to be advantageous for the management of fractures of the proximal humerus.