Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Experimental microdosimetric distributions of protons, from 19 to 65 MeV, using an A150-walled Low Pressure
M Takada1, H Yamaguchi, H Kitamura
1National Institute of Radiological Sciences, Chiba 263-8555, Japan. m_takada@nirs.go.jp
Abstract:
Microdosimetric distributions of protons, from 19 to 65 MeV, were measured using an A150-walled Low Pressure Proportional Counter at the cyclotron facility in the National Institute of Radiological Sciences. The measured distributions were found to be very different from the spectra assumed by the Continuous Slowing Down Approximation. The measurements consisted of determining the energy deposition by direct incident protons, by secondary electrons produced in the wall of the detector and by scattered protons at the wall. The secondary electron and the scattered proton distributions were studied using analytical functions and the MCNPX Monte Carlo code, respectively.