Estimating the Yield of Compounds on the TLC Plate via the Blue-LED Illumination Technique
Published on: October 6, 2022
A comparison of techniques for analysing 90Y-Zevalin thin-layer chromatography plates
Lesley M O'Brien1, Clinton C Waight, Alistair M Millar
1Radiopharmacy, The Royal Infirmary of Edinburgh, Edinburgh EH16 4SA, UK.
Objectives:
To simulate 90Y-Zevalin thin-layer chromatograms representing a range of radiochemical purities, to compare the radiochemical purities obtained with five techniques used to quantify 90Y on the plates and to measure the reproducibility of the five techniques at the minimum acceptable radiochemical purity of 95%.
Methods:
Yttrium-90 solutions were pipetted onto the origin and solvent front lines of thin-layer chromatography (TLC) plates to simulate radiochemical purities of 90%, 92%, 94%, 95%, 96%, 98% and 100%. Each plate was analysed using three TLC scanners (Bioscan AR2000, Bioscan Mini-scan and an instrument constructed in-house) and two cut-and-count techniques: one using a sodium iodide well detector and the other a liquid scintillation counter. The reproducibility of each technique was measured by analysing the 95% plate 10 times.
Results:
The radiochemical purities measured by the five techniques agreed well. The means of the seven results obtained with each agreed within 0.7%. The reproducibility of each technique was excellent. The coefficient of variation for 10 measurements was < or =0.3%. The signal to background ratios were satisfactory, ranging from 24 to 2.1 x 10(5).
Conclusion:
Each technique is suitable for analysing 90Y-Zevalin TLC plates.
Related Concept Videos
Chromatographic Methods: Terminology
Thin-Layer Chromatography (TLC): Overview
To begin the analysis, a mixture of compounds is spotted on the starting line on the TLC plate using a thin capillary. The bottom of the...
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Principles Of Column Chromatography
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...