[Evaluation of SMBG values using various instruments and clinical significance of forearm SMBG measurement]
Nayuta Mori1, Yuuko Oono, Yuuichi Murase
1College of Nutrition, Koshien University, Takarazuka 665-0006.
Abstract:
Whole blood or plasma glucose values were compared between those measured using various instruments for SMBG and automated analyses in samples (antecubital vein) from 20 approximately 70 years old obtained at cookie tests. A good correlation between values in whole blood SMBG and plasma automated analyses values greater than r = 0.94 was observed for instruments A, C, F, and H, and the per cent difference from the automated values was less than 5% for A, B, E and H. Per cent difference of SMBG values by I was -16% in whole blood and -9% in plasma, suggesting the possibility of measuring real values in whole blood. With plasma, a good correlation greater than r = 0.95 was noted for A, C, F and K, and the per cent difference less than 10% was noted for C, E, F and I. A relatively good correlation (r: 0.63 approximately 0.90) between forearm SMBG value and plasma automated values was noted with the % difference of the mean less than 11%. During cookie test, there is no significant difference between forearm SMBG values and antecubital plasma automated values. Values at finger tips are significantly greater by 5 approximately 20% over automated plasma values. In conclusion, whole blood values by most of the SMBG instruments are well correlated with plasma automated values, although some disagreement was noted. Values at forearm are not different from plasma val ues of antecubital vein, while at finger tip SMBG values showed higher levels. The measurement at forearm is therefore recommended, and in addition the pain is less.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement