Related Experiment Video
Updated: Aug 19, 2026

Clinical Imaging of Microwave Mammography
Published on: November 14, 2025
Molecular breast imaging: a new technique using technetium Tc 99m scintimammography to detect small tumors of the
Deborah J Rhodes1, Michael K O'Connor, Stephen W Phillips
1Department of Internal Medicine and Division of Preventive and Occupational Medicine, Mayo Clinic College of Medicine, Rochester, Minn 55905, USA. rhodes.deborah@mayo.edu
Objective:
To determine the sensitivity of molecular breast imaging (MBI) to detect small cancers of the breast.
Patients And Methods:
A cadmium-zinc-telluride gamma camera with a field of view of 20 x 20 cm was used. The detector elements were 2.5 x 2.5 mm. The gamma camera was mounted on a modified mammographic gantry. Between November 2001 and March 2004, we performed MBI on patients who were scheduled to undergo biopsy for a lesion suggestive of malignancy that was smaller than 2 cm on a mammogram. Patients were injected with 20 mCi of technetium Tc 99m sestamibi and underwent imaging immediately after injection. Using light pain-free compression, we obtained craniocaudal and mediolateral oblique views of each breast.
Results:
Of the 40 women included in the study, 26 had a total of 36 malignant lesions confirmed at surgery. Of these 36 lesions, 33 were detected by MBI (overall sensitivity, 92%). Of the 22 malignant lesions 1 cm or smaller in diameter, 19 were detected by MBI (sensitivity, 86%). Two patients had false-negative MBI results. Of the 14 malignant lesions larger than 1 cm in diameter, all were identified correctly by MBI. In 4 patients, MBI identified additional lesions not seen on mammography that were confirmed subsequently on magnetic resonance imaging and were true-positive cases at surgery. Three of these patients had lesions in the breast contralateral to the breast containing the initial mammographic finding suggestive of malignancy. Of 14 patients with no evidence of cancer at biopsy or surgery, 9 had true-negative (normal) scans and 5 had false-positive scans on MBI. False-positive results included benign fibroadenoma (2 patients), inflammatory fat necrosis (1 patient), benign breast parenchyma (1 patient), and complex sclerosing lesion (1 patient).
Conclusion:
This prototype gamma camera system for MBI reliably detects malignant breast lesions smaller than 2 cm. Furthermore, we obtained the highest sensitivity (86%) yet reported for the detection of lesions smaller than 1 cm. These results suggest an important role for MBI, particularly for women in whom the sensitivity of mammography is reduced by the density of the breast parenchyma.
Insights
Molecular breast imaging (MBI) demonstrates high sensitivity in detecting small breast cancers, including those 1 cm or smaller. This advanced imaging technique shows promise, especially for women with dense breast tissue where mammography may be less effective.
Area of Science:
- Radiology
- Nuclear Medicine
- Oncology
Background:
- Mammography sensitivity can be limited by dense breast parenchyma.
- Early detection of small breast cancers is crucial for improved patient outcomes.
Purpose of the Study:
- To evaluate the sensitivity of molecular breast imaging (MBI) for detecting small breast cancers (<2 cm).
Main Methods:
- A cadmium-zinc-telluride gamma camera was used to perform MBI on 40 women with suspicious lesions <2 cm.
- Patients received technetium Tc 99m sestamibi, and imaging was performed immediately.
- Standard craniocaudal and mediolateral oblique views were acquired with light compression.
Main Results:
- MBI detected 33 of 36 malignant lesions (92% overall sensitivity).
- For lesions ≤1 cm, MBI sensitivity was 86% (19 of 22 lesions).
- MBI identified additional lesions missed by mammography in 4 patients, including contralateral findings.
Conclusions:
- Molecular breast imaging reliably detects small malignant breast lesions (<2 cm).
- MBI achieved a reported high sensitivity of 86% for lesions ≤1 cm.
- MBI is particularly valuable for women with dense breasts, potentially improving cancer detection rates.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Magnetic Resonance Imaging