Comparison of indirect methods of measuring intra-abdominal pressure in children
Peter J Davis1, Satajyit Koottayi, Anna Taylor
1Paediatric Intensive Care Unit, Bristol Royal Hospital for Children, UK. Peter.Davis@ubht.swest.nhs.uk
Objective:
To determine the most accurate indirect method of measuring intra-abdominal pressure (IAP) in children.
Design And Setting:
Single-centre, prospective, clinical study in a 23-bed specialist paediatric intensive care unit in Australia.
Patients And Participants:
20 children admitted to paediatric intensive care with a peritoneal dialysis catheter in situ following congenital cardiac surgery.
Interventions:
IAP was measured directly via the peritoneal dialysis catheter and by intragastric manometry via an indwelling nasogastric tube, and by intravesical manometry via an indwelling transurethral urinary catheter, using volumes of 0, 1, 3 and 5 ml/kg body weight of sterile saline instilled into the bladder.
Measurements And Results:
Across the range of IAPs of 1-8 mmHg the Bland-Altman method for assessing agreement between two methods of clinical measurement showed bladder pressure measured via the urinary catheter with 1 ml/kg body weight of saline instilled to be the most accurate indirect measurement technique, tending to give pressures between 0.07 and 1.23 mmHg higher than the direct measurement (95% CI for bias). Measuring bladder pressure with either no saline instilled or more saline per kilogram body weight instilled was less accurate over the same range of pressures, as was measuring the gastric pressure.
Conclusions:
The most accurate indirect method of measuring IAP in children over the normal range of IAPs involves measuring bladder pressure via a transurethral urinary catheter with 1 ml/kg body weight of sterile saline instilled into the bladder.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Measurement of Blood Pressure