Related Experiment Video
Updated: Jun 8, 2026

Radiolabeling and Quantification of Cellular Levels of Phosphoinositides by High Performance Liquid Chromatography-coupled Flow Scintillation
Published on: January 6, 2016
Determination of bisphenol A using a flow injection inhibitory chemiluminescence method
Shuhao Wang1, Xinting Wei, Lingyun Du
1College of Environmental Science and Engineering, Donghua University, Shanghai 200051, People's Republic of China. shuhaowang@sohu.com
Abstract:
A new flow injection chemiluminescence (CL) method has been developed for the determination of bisphenol A (BPA), based on the inhibitory effect of BPA on the chemiluminescence reaction between luminol and potassium hexacyanoferrate. Under optimum conditions, the decrease in CL emission intensity was linear with BPA concentration in the range 8.0 x 10(-7)-1.2 x 10(-5) mol/L, and the detection limit was 3.1 x 10(-7) mol/L. The relative standard deviation (RSD) of 11 replicate measurements was 2.6% for 2.0 x 10(-6) mol/L BPA (n = 11). The sampling frequency was calculated to be ca. 120/h. This method has been successfully used to determine the content of BPA in aqueous solution of polycarbonate materials. A brief discussion on the possible chemiluminescence reaction mechanism is presented.