Related Experiment Video
Updated: Aug 19, 2026

Dithranol as a Matrix for Matrix Assisted Laser Desorption/Ionization Imaging on a Fourier Transform Ion Cyclotron Resonance Mass Spectrometer
Published on: November 26, 2013
Determination of dioscin in rat plasma by liquid chromatography-tandem mass spectrometry
Ke Li1, Yingwu Wang, Jingkai Gu
1Laboratory of Drug Metabolism and Pharmacokinetics, Shenyang Pharmaceutical University, Wenhua Road 103, Shenyang 110016, PR China.
Abstract:
A sensitive and specific high-performance liquid chromatography-tandem mass spectrometric (LC-MS/MS) assay for dioscin in rat plasma was developed. Ginsenoside Rh2 was employed as an internal standard. Dioscin is a naturally occurring saponin present in many traditional Chinese medicinal plants. Dioscin was determined after the acetonitrile-mediated plasma protein precipitation. The mobile phase consisted of acetonitrile:10 mmol/l aqueous ammonium acetate (95:5, v:v), which was pumped at 0.8 ml/min. The analytical column (100 mm x 4.6 mm i.d.) was packed with Hypersil ODS material (5 microm). The standard curve was linear from 1 to 100 ng/ml. The assay was specific, accurate (percentage deviations from nominal concentrations were <10%), precise and reproducible (within- and between-day coefficients of variation <10%). Dioscin in rat plasma was stable over three freeze-thaw cycles and at ambient temperatures for 24 h. The utility of the assay was demonstrated by determining dioscin plasma concentrations in five rats for 120 h following a single oral gavage dose of 90 mg/kg.