A biomechanical comparison of arthroscopic sliding and sliding-locking knots
Hussein A Elkousy1, Jon K Sekiya, Kathryne J Stabile
1Fondren Orthopedic Group, Sugar Land, Texas, USA.
Purpose:
To evaluate biomechanical characteristics of 3 arthroscopic sliding and sliding-locking knots and the square knot, the gold standard used in open surgery.
Type Of Study:
Biomechanical investigation.
Methods:
Four different knot types (Weston, square, Duncan loop, and Nicky's) were tested in 5 configurations in a closed-loop system on a materials testing device. Three of the 5 knots were backed up with 3 reverse half-hitches and alternating posts and all sliding and sliding-locking knots were tied using an arthroscopic technique. Twelve knots of each configuration were tested for loop security with a 7-N preload, and for knot security with load to failure at a strain rate of 1.25 mm/second and cyclic loading of a 30-N force for 50 cycles.
Results:
No knots subjected to the 7-N preload failed or slipped and all had similar elongation (0.1 +/- 0.1 mm) except Nicky's knot (0.3 +/- 0.2 mm). There was no significant difference in load at failure for the square knot (178 +/- 14 N), the Weston knot backed with 3 half-hitches (168 +/- 14 N), the Duncan loop (160 +/- 20 N), or Nicky's knot (148 +/- 13 N). Most knots with 3 half-hitches failed with rupture at the knot. Under cyclic loading, no knots failed and none elongated greater than an average of 0.3 mm.
Conclusions:
All knot configurations maintained high loop security. All sliding and sliding-locking knots backed with 3 half-hitches had load at failure comparable to the square knot. With cyclic load testing, all knots tested elongated minimally. Additionally, this study confirms that all knots, even the sliding-locking Weston knot, are best backed up with 3 half-hitches alternating posts and directions of the throws.
Clinical Relevance:
Sliding and sliding-locking knots are becoming increasingly popular among arthroscopic shoulder surgeons. This study provides a biomechanical basis for the clinical use of these arthroscopic knots and compares them with the gold standard, the open square knot.
Related Concept Videos
Structural Joints: Synovial Joints
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris group...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Structural Joints: Cartilaginous Joints
There are two types of cartilaginous joints:
Synchondrosis
A synchondrosis ("joined by cartilage") is a cartilaginous joint where bones are connected by hyaline cartilage. Synchondrosis may be temporary or...
Structural Joints: Fibrous Joints
Suture
All the bones of the skull, except for the mandible, are joined to each other by a fibrous joint called a suture. The fibrous connective tissue found at a suture strongly unites the adjacent skull bones and thus helps to protect the brain and form the face. In...