Measuring the real-ear to coupler difference transfer function with an insert earphone and a hearing instrument: are
1School of Psychological Sciences, Faculty of Medical and Human Sciences, University of Manchester, Oxford Road, Manchester, M13 9PL, United Kingdom.
Ear and Hearing
|February 5, 2005
Summary
The real-ear to coupler difference (RECD) varies between insert earphones and hearing instruments, impacting measurement accuracy. This study highlights potential discrepancies when using insert earphones for hearing instrument RECD testing.
Area of Science:
- Audiology
- Acoustics
- Hearing Instrument Technology
Background:
- The real-ear to coupler difference (RECD) is crucial for hearing instrument fitting and verification.
- Accurate RECD measurements ensure proper amplification and patient benefit.
- Variations in sound delivery systems can influence RECD outcomes.
Purpose of the Study:
- To compare RECD measurements obtained with an insert earphone versus two behind-the-ear hearing instruments.
- To assess the reliability and potential discrepancies in RECD measurements across different sound sources.
- To determine if insert earphones can accurately represent hearing instrument performance in RECD testing.
Main Methods:
- RECD was measured in 18 normal-hearing subjects using a probe-tube microphone system.
- Measurements were taken with a conventional ER-3A insert earphone and two hearing instrument models.
- RECD was calculated by subtracting 2-cc coupler (HA1 and HA2) responses from real-ear aided responses.
Main Results:
- The measurement procedures demonstrated high reliability with retest differences under 1 dB.
- Statistically significant differences (p < 0.05) were found between RECDs from insert earphones and hearing instruments.
- The largest RECD differences occurred with the HA2 coupler, reaching up to 11 dB between an insert earphone and a hearing instrument.
Conclusions:
- RECD is influenced by the acoustic impedance of the sound source, coupling system, and coupler.
- Differences in acoustic impedance between insert earphones and hearing instruments can lead to variable RECDs.
- RECD measurements using insert earphones may not always accurately reflect hearing instrument performance compared to coupler measurements.