Detachable coil for cerebral aneurysms: in vitro evaluation of magnetic field interactions, heating, and artifacts at 3T
Related Experiment Video
Updated: Aug 19, 2026

Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
Frank G Shellock1, Matthew Gounis, Ajay Wakhloo
1Keck School of Medicine, University of Southern California and Institute for Magnetic Resonance Safety, Education, and Research, 7511 McConnell Avenue, Los Angeles, CA 90045, USA. frank.shellock@gte.net
Abstract:
A detachable coil was assessed for magnetic field interactions, heating (body RF coil, SAR of 2.0-W/kg), and artifacts at 3-Tesla. The coil showed no magnetic field interactions and heating was negligible (+0.3 degrees C). Therefore, this coil is safe for a patient undergoing MR imaging at 3T or less. While artifacts may impair the ability to properly visualize anatomy in proximity to this implant, careful selection of imaging parameters can mitigate this problem.