Attenuation correction of myocardial SPECT studies using low resolution computed tomography images
Christine M Tonge1, Muthuvelu Manoharan, Richard S Lawson
1Department of Nuclear Medicine, Manchester Royal Infirmary, Oxford Rd, Manchester, M13 9WL, UK. Christine.tonge@cmmc.nhs.uk
Objectives:
Artefacts caused by tissue attenuation can create problems in the interpretation of myocardial perfusion studies. This study aimed to determine if attenuation correction using the 'Hawkeye' low-resolution X-ray computed tomography facility attached to a GE Millennium VG gamma camera reduced attenuation artefacts in our patient group. This technique offers potential advantages over isotope transmission methods of attenuation correction because of the lack of noise in the transmission images.
Methods:
Ninety-four patients (64 males, 30 females) underwent stress/rest myocardial perfusion imaging using (99m)Tc tetrofosmin (188 studies). Semi-quantitative analysis of the attenuation corrected and non-corrected perfusion images was carried out using proprietary software (Emory ECToolbox).
Results:
No significant differences between attenuation corrected and non-corrected images were seen in the septal, lateral or apical segments. Attenuation correction produced a significant reduction in the defect scores in the inferior segment indicating the presence of attenuation artefacts. This was more marked in male patients. The incidence of attenuation artefacts in the inferior segment increased with weight although patients of normal weight also showed attenuation artefacts. The changes in the anterior segment were more difficult to interpret particularly in females, with attenuation correction increasing the defect score in some patients. Minor degrees of mal-registration (even of 1 pixel) between emission and transmission images increased the likelihood of creating a defect.
Conclusion:
Attenuation correction using Hawkeye is likely to improve diagnostic accuracy in men, but is less likely to be useful in women.
More Related Videos
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Acute Coronary Syndrome III: Diagnostic Studies