Supplemental inhaled gases alter tidal volume delivery and measurement
David S Cooper1, Steven M Schwartz, Jenni L Raake
1Division of Cardiology, Cincinnati Children's Hospital, OH, USA.
Objective:
Supplemental inspired nitrogen (N(2)) or carbon dioxide (CO(2)) is commonly used to balance pulmonary blood flow in patients with single-ventricle physiology. The objective of this study was to assess if supplemental inspired gas alters delivery or measurement of tidal volume (V(T)) by a ventilator.
Design:
Prospective, experimental study.
Setting:
Respiratory Care Laboratory, Cincinnati Children's Hospital.
Interventions:
Using a test lung, expired V(T) measurements from Servo 300 ventilators were compared with actual delivered V(T) (true V(T)) at baseline and during supplemental N(2) or CO(2) administration to mimic clinical use in single-ventricle patients. At compliance settings simulating normal and compromised lung function, true V(T) was determined by the test lung and inline Pneumotach. True and measured V(T) were compared by repeated-measures analysis of variance with significance defined as p < .05.
Measurements And Main Results:
With normal lung compliance, supplemental gas administration increases both true and measured V(T), and expired V(T) measurements remain accurate. With poor lung compliance, supplemental gas flow disproportionately affects V(T) measurement. Poor lung compliance reduces true V(T) markedly (p < .001), causing a large discrepancy between true and measured V(T). Supplemental gas administration amplifies this discrepancy because the additional gas flow in the circuit erroneously augments expired V(T) measurements by the ventilator (p < .001). The discrepancy is greatest with higher-set V(T) and greater supplemental gas flow.
Conclusions:
The addition of supplemental inspired gas directly into the ventilator circuit can alter tidal volume delivery or measurement by a ventilator. The extent and magnitude of the alterations are determined by lung compliance. Variable effects of supplemental gas administration may confound ventilator management of patients with single-ventricle physiology.
More Related Videos
08:44Dual Test Gas Pulmonary Diffusing Capacity Measurement During Exercise in Humans Using the Single-Breath Method
Published on: February 2, 2024
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Physical Principles Governing Gas Exchange
Gas Laws Governing Respiration
The behavior of gases is guided by Dalton's Law of partial pressures and Henry's Law.
Dalton's Law asserts that the total pressure exerted by...
Inhaled Medications
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...