An integrated dual ultramicroelectrode with lower solution resistance applied in ultrafast cyclic voltammetry
Zhiyong Guo1, Xiaohua Jiang, Xiangqin Lin
1Department of Chemistry, University of Science and Technology of China, Hefei 230026, P R. China.
Abstract:
A method for calculating the solution resistance of an integrated dual ultramicroelectrode was introduced, and then tested by experiments using dummy cells. Then, with the reduction of anthracene on a gold ultramicroelectrode in a 0.1 M tetraethylammonium tetrafluoroborate/acetonitrile solution as the test system, it could be found that the solution resistance of this integrated dual ultramicroelectrode was much lower than that of the two-electrode system conventionally used. Thus, the compensation level could be improved up to nearly 100% at a scan rate of 1.34 MV s(-1), while the latter could not. This showed that the integrated dual ultramicroelectrode was more suitable for acting as the electrode system in ultrafast cyclic voltammetry, especially in a high-resistance solution.
Related Concept Videos
Interfacial Electrochemical Methods: Overview
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetric Techniques: Pulse Voltammetry
Voltammetric Techniques: Cyclic Voltammetry
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...