Chest radiography with a digital flat-panel detector: experimental receiver operating characteristic analysis.

Stephan Metz1, Petra Damoser, Regina Hollweck

  • 1Department of Diagnostic Radiology and Institute of Medical Statistics and Epidemiology, Technische Universität München, Ismaninger Str 22, 81675 Munich, Germany. smetz@roe.med.tu-muenchen.de

Radiology
|March 1, 2005
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Whole body CT attenuation and volume charts from routine clinical scans via LLM report filtering.

NPJ digital medicine·2026
Same author

Autologous minced cartilage implantation for treatment of focal cartilage defects of the glenohumeral joint results in good clinical and radiological outcomes after a minimum 2-year follow-up.

Journal of experimental orthopaedics·2026
Same author

International survey in four European countries on diagnostic reference levels based on clinical indications in computed tomography.

Insights into imaging·2026
Same author

Nodular fasciitis in unusual (and usual) locations: lessons learned from a challenging diagnosis.

Pathology·2026
Same author

Correction: Requirements for Physico-Technical Quality Assurance in the Framework of Early Detection of Lung Cancer.

RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin·2026
Same author

Body Charts from CT Segmentations across the Adult Lifespan: Large-scale Cross-sectional and Longitudinal Analyses.

Radiology. Artificial intelligence·2025

Reducing radiation dose in chest radiography with flat-panel detectors (FPDs) can decrease diagnostic performance, especially for overall lesion detection. However, lung field lesion detectability remains largely unaffected even with a 50% dose reduction.

Area of Science:

  • Medical Imaging
  • Radiography
  • Diagnostic Performance

Background:

  • Flat-panel detectors (FPDs) are widely used in digital radiography.
  • Optimizing radiation dose while maintaining diagnostic quality is crucial in chest radiography.

Purpose of the Study:

  • To assess how varying detector radiation doses and peak kilovoltage (kVp) settings impact diagnostic performance and radiation dose in posteroanterior (PA) chest radiography using an FPD.

Main Methods:

  • PA chest radiographs of a phantom were acquired using an FPD at different detector doses (2.50, 1.56, 1.25 microGy) and kVp settings (100, 120, 140 kVp).
  • Simulated lesions were superimposed, and four radiologists evaluated image quality.
  • Receiver operating characteristic (ROC) analysis was performed, and entrance surface dose and effective dose were measured/calculated.

Related Experiment Videos

Main Results:

  • Lower detector doses significantly reduced overall lesion detection performance (P < .05).
  • However, lesion detectability within pulmonary areas was not significantly affected by dose reduction.
  • 120 kVp offered the best balance between diagnostic performance and effective dose.

Conclusions:

  • Standard PA chest radiography should maintain a detector dose of 2.50 microGy (speed, 400) with 120 kVp for optimal diagnostic performance.
  • Significant loss in diagnostic performance was not observed for lung field imaging even with a 50% radiation dose reduction.