The impact of closed endotracheal suctioning systems on mechanical ventilator performance
Ashraf El Masry1, Purris F Williams, Daniel W Chipman
1Department of Anesthesia and Critical Care, Massachusetts General Hospital, Boston 02114, USA.
Background:
Closed endotracheal suctioning during mechanical ventilation is increasingly used, but its impact on ventilator function has not been fully studied.
Methods:
We evaluated the impact of closed suctioning with 11 critical-care ventilators, during assisted ventilation in pressure-support mode, pressure-assist/control mode, volume-assist/control mode, and during continuous positive airway pressure, with 2 suctioning pressures (-120 mm Hg and approximately -200 mm Hg), and with 2 tidal volumes (450 mL and 900 mL). We continuously measured airway pressure, flow at the airway, and pressure distal to the catheter tip, before, during, and after a single 15-second period of continuous suctioning.
Results:
No ventilator malfunctioned as a result of the closed suctioning. During suctioning, end-expiratory pressure markedly decreased in all modes, and peak flow increased in all modes except volume-assist/control (p < 0.001). Respiratory rate increased during suctioning in pressure- and volume-assist/control (p < 0.001) but not during pressure support or continuous positive airway pressure. Gas delivery was most altered during volume-assist/control with the smaller tidal volume (p < 0.05) and least altered during pressure-assist/control with the larger tidal volume.
Conclusion:
There are large differences between the ventilators evaluated (p < 0.001). Closed suctioning does not cause mechanical ventilator malfunction. Upon removal of the suction catheter, these ventilators resumed their pre-suctioning-procedure gas delivery within 2 breaths, and, during all the tested modes, all the ventilators maintained gas delivery. However, closed suctioning can decrease end-expiratory pressure during suctioning.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Tracheostomy Suctioning II: Procedure
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)