[Determination of dimethyl ether correction factors in gas chromatography with TCD and FID]
1Dalian Institute of Chemical Physics, the Chinese Academy of Sciences, Dalian, 116012.
Se Pu = Chinese Journal of Chromatography
|May 1, 1997
Abstract:
Dimethyl ether (DME) correction factors in gas chromatography with thermal conductivity detector (TCD) and flame ionization detector (FID) by using H2 as carrier gas were determined in this work. The homemade DME gas was quantitatively absorbed in ice-cold water. With ethanol as standard, the aqueous mixture was injected into a gas chromatograph, equipped with serially-connected TCD and FID. The weight correction factors of DME based on methanol were 0.86 and 0.55 for TCD and FID respectively. The result for TCD was also confirmed by calculation based on the stoichiometrical transformation of methanol into DME in reaction gas chromatography.