On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
Published on: August 5, 2016
[Capillary gas chromatographic analysis of micro impurities in acrylonitrile]
Abstract:
A megabore SCOT series column of silicone OV-101 and OV-17 for the analysis of micro amount of impurities in acrylonitrile is presented. By using the columns of 50m x 0.53mm i.d. OV-101 and 25m x 0. 53mm i.d. OV-17, with column temperature of 75 degrees C and N2 as carrier gas, u=8cm/s, the main impurities, such as acrolein, acetone and acetonitrile may be well separated from acrylonitrile in 26 min. The results indicate that the series column possesses of high separation efficiency and long column life. The error in quantitative analysis was less than 4%.
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Types of Detectors-II