Level of disagreement between Proview phosphene tonometer and Goldmann applanation tonometer intraocular pressure
Sushma Rai1, Marlene R Moster, Muge Kesen
1Wills Eye Hospital, William and Anna Goldberg Glaucoma Service and Research Laboratory, Philadelphia, Pennsylvania 19107-5109, USA.
Purpose:
To evaluate the agreement of intraocular pressure (IOP) readings obtained with the Proview phosphene tonometer and those obtained by Goldmann applanation tonometry as well as the effect of regular use of the Proview on patients' anxiety about their glaucoma.
Methods:
One hundred thirty-five consecutive patients with glaucoma, 35 designated as controls, were enrolled in a 10-month randomized prospective clinical trial. The study patients but not the controls used the Proview outside the office. At office visits IOP was measured by an ophthalmologist with the Goldmann applanation tonometer (GAT) as well as by examiners and patients using the Proview. Patients in the treatment group (n = 100) were asked to measure and record their IOP at home as well. Using ANOVA, we examined the relationship between demographic variables and the level of agreement between the Proview readings and those measured by GAT at the final office visit. All participants also were asked to complete a questionnaire regarding their anxiety about their IOP at the baseline and final office visits. The primary outcome measure was the level of agreement of the Proview measurements with those obtained by GAT. Secondary outcome measures included patients' anxiety about their glaucoma.
Results:
The absolute mean difference between GAT and Proview readings at the final visit was 3.5 +/- 2.9 mm Hg (median, 2.8 mm Hg). The treatment group reported significantly less anxiety about their glaucoma after 4 to 6 weeks of using the Proview (P = .024).
Conclusions:
There was considerable discrepancy between Proview and GAT readings. However, regular use of the Proview tonometer significantly reduced patients' anxiety about their glaucoma.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Glaucoma: Overview
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...