Calculation of magnetic field-induced current densities for humans from EAS countertop activation/deactivation
1Department of Electrical and Computer Engineering, University of Utah, Salt Lake City, UT 84112, USA.
Abstract:
Compliance testing of electronic article surveillance (EAS) devices requires that induced current densities in central nervous system (CNS) tissues, i.e. brain and the spinal cord, be less than the prescribed safety limits. Even though ferromagnetic cores are mostly used for activation/deactivation of embedded magnetic tags, assumed equivalent air-core coils with guessed increased number of ampere turns have always been used to calculate the magnetic fields for the proximal region to which a customer is exposed. We show that at low frequencies up to several kilohertz, duality of electric and magnetic circuits may be exploited such that the shaped high reluctance core is modelled as though it was a higher conductivity electric circuit of the corresponding shape. The proposed procedure is tested by examples of two magnetic cores typical of countertop activation/deactivation devices. The equivalent exposure magnetic fields obtained from the dual electric fields are shown to be in excellent agreement (within +/-5%) with those measured for these ferromagnetic EAS devices. The previously proposed impedance method is then used to calculate the induced current densities for a 1.974 x 1.974 x 2.93 mm resolution anatomic model of a human. For the two considered EAS systems using excitation currents of 5000 A turns at 200 Hz, the maximum 1 cm2 area-averaged induced current densities in the CNS tissues are calculated and found to be less than the ICNIRP safety limits.
Related Concept Videos
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Magnetic Force On A Current-Carrying Conductor
Consider a compass placed near a current-carrying wire. The wire experiences a force that aligns the needle of the compass tangentially around the wire. Thus, the current-carrying wire produces concentric circular loops of magnetic field. The magnetic field generated by a wire can be...
Magnetic Field Of A Current Loop
Faraday's Law
Induced Electric Fields: Applications
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...