Safety of magnetic resonance imaging of stapes prostheses
1Barrow Neurologic Institute, Phoenix, Arizona, USA. m.syms@arizonaear.com
Objective:
Assess the safety of performing magnetic resonance imaging (MRI) on patients with stapes prostheses.
Study Design:
Survey and animal model.
Methods:
A survey regarding implant usage, MRI procedures, and adverse outcomes after MRI in patients previously undergoing stapes procedures. Guinea pigs implanted with ferromagnetic 17 to 4 stainless steel, 316L nonferromagnetic stainless steel, titanium, and fluoroplastic stapes prostheses underwent a MRI in a 4.7 Tesla MR system.
Results:
: Three adverse outcomes were reported on the clinical survey. One adverse event occurred during an MRI performed on a recalled ferromagnetic prosthesis. The other two adverse events were probably not secondary to MRI exposure. No damage or inflammation was observed in the region of the oval window or vestibule of implanted guinea pigs exposed to a 4.7 Tesla MR system.
Conclusions:
The combination of prior studies, the clinical survey, and the absence of histopathologic evidence of damage in the guinea pigs is compelling evidence that MRI for patients with stapes prostheses is safe. Implanting physicians should feel comfortable clearing a patient for a MRI in a 1.5 Tesla or 3.0 Tesla MRI. It is imperative for the physician to qualify the field strength when clearing a patient to undergo a MRI.
More Related Videos
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...