Related Experiment Video
Updated: Aug 19, 2026

Remote Laboratory Management: Respiratory Virus Diagnostics
Published on: April 6, 2019
Residual formaldehyde on plastic materials and medical equipment following low-temperature steam and formaldehyde
K Kanemitsu1, H Kunishima, T Saga
1Department of Molecular Diagnostics, Tohoku University Graduate School of Medicine, 1-1 Seiryo-machi, Aoba-ku, Sendai, Miyagi 980-8574, Japan. kane2@mail.tains.tohoku.ac.jp
Abstract:
We measured the amount of residual formaldehyde on 16 plastic materials and five medical devices following low-temperature steam and formaldehyde (LTSF) sterilization, based on the European Standard EN14180. The amounts of formaldehyde residue on the plastic materials were compared with that on a filter paper of similar dimensions. The amount of residual formaldehyde on polyamide 6, polyurethane, natural rubber and polyacetal was higher (21.9, 15.2, 3.0 and 2.1 times, respectively) than that on the filter paper. The amount of formaldehyde recovered from a breathing circuit, anaesthesia circuit, oxygen tubing, airway tube and tweezers was 260, 240, 594, 56 and 0 microg, respectively, following LTSF sterilization. Our results emphasize the need to verify the main material composing the medical equipment before LTSF sterilization, as the amount of formaldehyde retrieved following sterilization varies according to the material used for construction.
More Related Videos
14:17The Portable Chemical Sterilizer (PCS), D-FENS, and D-FEND ALL: Novel Chlorine Dioxide Decontamination Technologies for the Military
Published on: June 29, 2014
06:47Effects of Exposure of Formaldehyde to a Rat Model of Atopic Dermatitis Induced by Neonatal Capsaicin Treatment
Published on: September 27, 2017
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Temperature
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Chemical Agents for Microbial Control
Physical Methods for Controlling Microbial Growth: Radiation and Filtration