Erroneous reporting of coagulase-negative Staphylococci as Kocuria spp. by the Vitek 2 system
R Ben-Ami1, S Navon-Venezia, D Schwartz
1Infectious Diseases Unit, Tel-Aviv Sourasky Medical Center, Tel-Aviv 64239, Israel.
Abstract:
Misidentification of coagulase-negative staphylococci (CoNS) may delay appropriate treatment. We investigated 20 clinical isolates identified as Kocuria spp. by the Vitek 2 system. All were identified as CoNS by 16S rRNA gene sequencing (18 Staphylococcus epidermidis, 1 Staphylococcus haemolyticus). Four Kocuria isolates were shown to be identical to CoNS from the same patient by pulsed-field gel electrophoresis. Isolates identified by Vitek 2 as Kocuria most likely represent misidentified CoNS, and if clinically indicated, should be investigated further by genomic methods.