Minimal Erythema Dose (MED) Testing
Published on: May 28, 2013
Ultraviolet A dosimetry in photopatch test centres in Europe
1The Photobiology Unit, University of Dundee, Ninewells Hospital and Medical School, Dundee DD1 9SY, Scotland, UK. h.moseley@dundee.ac.uk
Objective:
The aim of the study was to investigate the state of ultraviolet (UV) A dosimetry in photopatch testing centres in Europe.
Background:
It has been acknowledged that a widespread disparity exists in the way photopatch testing is carried out in different countries throughout Europe. This prompted the formation of a European taskforce that developed a consensus methodology. One issue that was not addressed directly was that of UVA dosimetry, which clearly impacts on the consistency of the clinical results.
Study Design:
Ten UV meters from photopatch test centres throughout Europe were set up at a reproducible position from a photopatch test light source containing UVA fluorescent lamps and the irradiances were noted from each meter. One meter served as a reference meter with calibration carried out in an optical laboratory traceable to the National Physical Laboratory.
Results:
The mean of all the measurements was 5.74 mW/cm(2) with a SD of 0.36 mW/cm(2). Expanded uncertainty at 95% confidence level was 12%.
Conclusions:
It is desirable that meter calibration should have an accuracy of the order of +/-10%. The variation in recorded values among the 10 meters investigated was within acceptable limits. The meter must be calibrated using a similar light source to that used in the phototest equipment, in this case a UVA fluorescence lamp.