A dwell position verification method for high dose rate brachytherapy
Lizhong Liu1, Satish C Prasad, Daniel A Bassano
1Department of Radiation Oncology, SUNY Upstate Medical University, 750 East Adams Street, Syracuse, New York 13210, USA. liul@upstate.edu
Abstract:
Misplacement of dwell positions is a potential source of misadministration in high dose rate brachytherapy. In this work we present a dwell position verification method using fluoroscopic images. A mobile C-arm fluoroscopic machine is used to take a snapshot of the treatment machine's check cable as it reaches the most distal dwell position. This fluoroscopic image is displayed side-by-side with a treatment planning image on a dual monitor relay station at the HDR treatment console. Any discrepancy between the check cable's position on the verification image and the intended dwell position on the planning image can be identified immediately, thus avoiding the possibility of treating the wrong target volume.